Temperature Controller 
Our Temperature Controllers are designed to work with different heating elements and a variety of optional accessories including Dry Bath, Water Bath, BLM Cell Holder and Extruder Warmer.  Two modifications are available.  TC-02A can be connected to high power AC heaters (lamps, heater fans, water immersed heaters etc.) for large volume heating, or to compact heater assemblies for temperature control of fluorometer cells, or microscope stages.  TC-12 version is similar but smaller and have lower power, no AC and no recorder output. Each unit comes with one temperature sensors attached to 3 feet (90 cm) cables.  Second sensor assembly (Part # AD592) is optonal and need to be purchased separately.
 
SPECIFICATIONS
TC-12 TC-02A
Temperature setting range, oC 20-80 20-80
Temperature measurements range, oC  -10 to +100  -10 to +100
Resolution,  oC 0.1 0.1
Accuracy, oC +/-1 +/-1
Recorder output No BNC, 1 mV/oC
Number of sensor inputs 2 2
Sensor dimensions LxD (mm) 4x5 4x5
Controller dimensions, HxWxL (sm.) 9x16x12 (3.5"x6"x5") 9x21x15  (3.5"x8"x6")
Output power 
(2.1/5.5 mm DC connector;  standard AC socket)
12W, 24V DC  24W, 24V DC ; 
300W, 120V AC



 
 
Heater assemblies
Part #
Peak power with 
TC-12 or TC-02A
(Watt)
Mean power (Watt)
Resistance (Ohm)
Body Size (mm)
Comments
HA-5
6
5
100
 10x10x23
Ceramic
HA-5M
6
15
100
8x11x3
TO-126
HA-10
12
10
50
 10x10x50
Ceramic
HA-10M
12
30
50
10x16x3
TO-220
HA-20
23
20
25
 10x10x50
Twin Ceramic
HA-20M
23
60
25
10x16x3
Twin TO-220

All heater assemblies are attached to 4 feet (120 cm) cable with 2.1/5.5 mm connector that match TC-12/TC-02A output.  Heated surfaces are electrically non conductive and can be mounted directly to metal surfaces using enclosed hardware.  Ceramic heaters can be used at maximum power without heat sink (but with sufficient air cooling) and good for temperatures up to +150oC.  TO-XXX heating elements need good contact with metal heat sink (use of silicone heat sink compound is recommended).  They can be irreversibly heat damaged if used unmounted.  It is not recommended to use them for heating to temperatures above +80oC.



Dry Bath
This compact unit is designed to work with  TC-02A or other temperature controllers that have 12-24 V output.  Allow keeping small number of microcentrifuge tubes (0.5-2 ml size) above room temperature near your workspace.  Two ceramic heating elements attached to the bottom of massive aluminum alloy heating block provide even heat distribution to up to 15-16 vials.  (One-two well should be reserved for feedback temperature sensor and thermometer).
Part # DB-1

Specifications:
Max. Temperature     110 oC
Max. Power              25 Watt
Heater resistance       25 Ohm
Dimensions               90x125x50 mm (3.5"x5"x2")



Water Bath
Micro Water Bath is the another option for keeping samples close to the workplace and at required temperature.  This attachment has water immersed heater and a dry well for temperature sensor.  Fill it with water, place on top of a magnetic stirrer (not included) and connect to TC-02A controller, and you have compact, simple and reliable bath for different kind of lab tubes or small flask.
Part # WB-1
 
 
 
 

Specifications:
Max. Temperature     60 oC
Max. Power              25 Watt
Heater resistance       25 Ohm
Usable space             75x75x60 mm (3"x3"x2.5")
Overall Dimensions    90x125x75 mm (3.5"x5"x3")



OH-xx Series Objective Heaters provide a simple and effective method for reducing thermal gradient between objective and specimen.  Designed to work with TC-12 temperature controller (TC-02 can be used as well). Objective Heater represent collar that attaches to microscope objective. This collar incorporates a resistor block  and holder for AD592 feedback temperature sensor. When connected to TC-12 controller, electronic controls  maintain the objective at a constant pre set temperature.
 

To determine which heater is right for your objective:

1. Measure the length of the objective cilindrical part were heater to be mounted.  For standard heaters it should be 12 mm or more. If it is less than 12 mm you need custom made heater.
2. Measure the body diameter of the objective near the lens end.
With these measurements choose the objective heater from the table below. Custom designs are also available.
 
OH-20
OH-26
OH-35
Fit Diameters (mm)
18-22
24-28
32-38
Collar width (mm)
12
12
12
Temperature range
Ambient to +40°C 
Ambient to +45°C 
Ambient to +45°C 