With these devices you can put images from two video cameras on the single video monitor. This may be extremely useful when you need to find correlation between two different processes. For example, if one camera images cells and another one is focused on pH meter display, you will be able to detect pH dependent changes in cell morphology with 1/30 second precision. VSS-1 unit split screen either vertically or horizontally and allow to change area allocated to each camera and adjust camera gain. VSS-2 unit inserts variable size rectangle of camera 2 image into camera 1 field. Both units work with any black and white analog cameras but one camera must have video synchronization input (usually standard BNC connector labeled "Gen. Lock" or "Sync.")
Part # ADA-1210
Specifications
Analog input
Input voltage
±10 Volts (V) max
Input impedance
122 kOhm
Number of channels
4, software selectable
Input ranges
±20 V, ±10 V, ±5 V, ±4 V, ±2.5 V,
(Software selectable)
±2.0 V, ±1.25 V, ±1.0 V
Sampling rate
250 samples per second (S/s) typ
(PC-dependent)
Continuous scan
50 kilo samples per second (kS/s)
Resolution
12 bits, no missing codes
Analog output
Resolution
12-bits, 1 in 4096
Output range
0 - 4.096 V, 1 mV per LSB.
Number of channels
2
Sampling rate
250 S/s single channel typical,
(PC dependent)
Single channel, continuous scan
10 kS/s
Dual channel, continuous scan,
5 kS/s
Digital I/O specifications
Digital type
CMOS
Number of I/O
4
Input high voltage
2.0 V min, 5.5 V max
Input low voltage
0.8 V max, 0.5 V min
Picoamp-300B
This amplifier is a modification
of PA-300 model.
In designet to work with new HS-2C
headstage (shown on top of the unit) for cell electrophysiology applications.
This headstage has AxonTM
type threaded connector for stable and easy microelectrode mounting, 1
mm diameter gold plated connector for ground electrode and 2-56 screw for
system ground. Headstage connects to the amplifier with our standard
RJ-45 connector and backward compatible with PA-300 and even with BBA-02
units. However, "B" model amplifier has added Zap circuit which generate
short pulses of membrane breaking voltage and that is quite useful for
converting cell "patch" to "whole cell" configuration. Additionally, offset
and RC compensation ranges are increased to be compartible with glass microelectrodes.
This makes PA-300B
truly universal for both basic cell electrophysiology applications and
for BLM research. Once you decided to get to the basics of cellular
phenomena using artificial lipid bilayers, you only need to replace HS-2C
headstage with HS-2
or HS-2A, no
recalibration required.
Main Unit Part # PA300B;
Headstages Part ## HS-2, HS-2A orHS-2C
SPECIFICATIONS:
| Potential on membrane | 0 to +/- 200 mV ; 0 to +/- 500 mV |
| Offset potential comp. | >= +/- 50 mV |
| Gain Multiplier | x 2, x 5 |
| Filter | 40 dB, low pass, Bessel. 1 kHz; 5 kHz, |
| Electrode series
resistance compensation |
0-50 kOhm |
| Max. output voltage | +/- 10 V |
| Zap pulse duration | 2-50 ms |
| Display | VBLM/VOUT; DC/RMS, 0-200 mV/ 0-2V |
| 50/60 Hz test signals | Selectable 5 mV rectangular; ~ 10 mV triangular |
| Ultra low frequency test signals | Selectable 0 to +/- 200; 0 to +/- 500; rectangular/triangular; 1/2/4 min. period |
| Dimensions | 42x9x20 cm (17"x3.5"x8") |
| Power | AC 115/230 V, 6 W. |
|
|
|
|
|
| Gain, mV/pA |
|
|
|
| Input bias current, pA |
|
|
|
| Noise, pA * |
|
@0.1/1 pA/mV |
|
| Frequency range, kHz ** |
|
@1/0.1 pA/mV |
|
| Membrane capacitance
|
|
|
|
| Zap pulse amplitude (mV) |
|
|
|
| Dimensions, cm |
|
|
|
* Peak to peak, with 1 kHz low pass filter. Measured with open input.
** at 3 dB nonlinearity.